The Deep Water Gas Analyzer combines membrane separation technology with Los Gatos Research’s cavity enhanced laser absorption spectroscopy technology to provide in situ measurements of methane and carbon dioxide at depths to 2500 meters. Designed for applications such as ocean acidification, carbon sequestration, methane hydrate and hydrothermal vent studies, these novel analyzers provide real-time measurements of dissolved gases in oceans and lakes.
Performance Specifications
Gases Measured (one configuration per instrument):
Configuration 1: Methane concentration
Configuration 2: Carbon dioxide concentration
Configuration 3: δ13C in methane (δ13CH4)
Precision (1σ):
Configuration 1: [CH4]: 1 ppbv
Configuration 2: [CO2]: 1 ppmv
Configuration 3: δ13CH4: 1‰
[CH4]: 1 ppmv
Measurement Range (meets all specs):
Configuration 1: [CH4]: 1– 500 ppmv
Configuration 2: [CO2]: 300 – 25000 ppmv
Configuration 3: [CH4]: 400 – 20000 ppmv
Operational Range
(external calibration may be required):
Configuration 1: [CH4]: 0.5 – 1000 ppmv
Configuration 2: [CO2]: 150 – 50000 ppmv
Configuration 3: [CH4]: 200 – 40000 ppmv
Measurement Rate:
5 minutes
Maximum Depth:
2500 meters
Sampling conditions:
Sample Temperature: 0 – 50 °C
Operating Temperature: 5 – 45 °C
Outputs:
Digital (RS232), Ethernet
Power Requirements:
115/230 VAC, 50/60 Hz
80 watts
Dimensions:
20 cm diameter
122 cm length
Weight:
100 pounds (45 kg)
|